REFERENCE AUTOMATIC LIQUID DENSITY MATER MD
Purpose of the development:
Improvement of accuracy and metrological reliability of results of liquid density measurement.
Recommended application field:
Different industries of national economy.
The development stage readiness:
Tested experimental exploitation order
Description of the development:
() The device consists of a measuring chamber with a placed inside float with a constant magnet, a measuring solenoid located outside the chamber, block of analog-digital transformation, analog managing block, digital managing and computing blocks, a float position gauge containing an induction coil. The density meter principle of operation is based on equilibration of the expulsive force and force of interaction with the solenoid magnetic field. With the aid of the built-in platinum thermometer of resistance, the temperature is measured simultaneously with the liquid density.
Possibility of transfer abroad:
Licence's sale Technological document's sale Creation of joint enterprise
Photo
Country
Russia
For additional information turn to: E-mail: gal@uintei.kiev.ua
|