DEVICE FOR COLD PLASMA STERILIZATION OF MEDICAL PRODUCTS WITH DC GLOW DISCHARGE (SGR-100)
Purpose of the development:
PDS plasma sterilization system 100 is designed for cold sterilization of medical instruments containing temperature-sensitive components . These tools can take place , which is difficult to diffusion , namely , the inner surface of the slit or a long narrow tube. Using plasma sterilizer PDS 100 not only provide a solution to environmental problems , but also lead to a significant increase in economic efficiency by reducing the duration of the sterilization cycle sterilization.
Plasma sterilization technology approved by the Ministry of Health of Ukraine .
Recommended application field:
Medicine.
Technical characteristic:
Main technical characteristics:
Overall dimensions 810 x 620 x 1510 mm
Camera depth: 650 mm, diameter 450 mm
Material stainless steel chamber 12X18H10T
Weight 275 kg
Operating conditions T = 10 - 43 ° C;
relative humidity: 80 %
atmospheric pressure: 600-800 Top
Average power consumption - 500 Watts Peak - 1100 W
Management system ( fully automatic ) embedded microcontroller based processor 80c31, sound alarm
Maximum temperature of products to be sterilized 60 ° C
Turnaround times ( from downloading to discharge ) 15-45 min.
Time for complete sterilization of the exposed surfaces of packaged items : 5 min . ;
product with the blind holes of diameter 3 mm and length of 650 mm for 30 minutes.
Advantages over analogues:
advantages:
Proved experimentally that this plasma sterilizer exhibits high efficiency sterilization of up to 100% compared with conventional sterilizers (12/88 for ethylene oxide) as well as with other types of plasma sterilizers (Sterrad 100, Advanced Sterilization Products; Plazlyte Sterilization System, model ABT 1.0 , Abtox Inc.).
The development stage readiness:
Ready for application
Description of the development:
() Job plasma sterilizer is based on the use of a dc glow discharge , ensuring the creation of a dense plasma and visokoodnorodnoy directly in the space , in which there are products that are sterilized . This discharge plasma is more efficient for the rapid sterilization of medical instruments , especially products with long passages or holes. The sterilization process comprises repeated cycles of treatment consists of two stages - the "steam " and " plasma ." During the first phase, the chemical vapor is injected into the pre- evacuated sterilizer chamber . A special system of injection and evaporation provided a very effective steam penetration into difficult places and packages of processed products . After the " vapor " phase again evacuated chamber and ignited a dc glow discharge . Medium which is used for plasma generation , a mixture of the vapor of hydrogen peroxide and water, and molecular oxygen generated by the decomposition of hydrogen peroxide.
corresponds technical description
Ready for implementation
Possibility of transfer abroad:
Licence's sale Technological document's sale Creation of joint enterprise
Photo
Country
Ukraine
For additional information turn to: E-mail: gal@uintei.kiev.ua
|